Climax Technology Ventures, Inc.
Specification for Model 9705 Power Supply
Operation
At power "ON", the strike pulse is delayed for a 30 second filament warmup at the high voltage setting (10VDC*). A 700 V strike pulse is applied to the Anode voltage. If the lamp ignites, the current is regulated at the Anode current setting (300*+-0.03 mA),The filament is kept at the high voltage level for another 30 seconds to stabilize the current. Then the filament is switched to the lower setting (7VDC*). If the tube does not ignite or the curent ceases, the Autostart cycle begins again.The factory voltage and current settings (*) listed above can be user adjusted.
Input/Output SpecsAC Input VoltageSafety and Warnings
115 VAC rms Nominal (90-132 VACrms) with 115V selection 230 VAC rms Nominal (180-264 VACrms) with 230 V selection AC Power Draw
100VA maximum Frequency
48 to 63 Hz Power Factor
0.89 Fuses (External Access)
Two type 3AG 2 A at 115 VAC input Two type 3AG 1 A at 230 VAC input Filament Voltage
High ("Start" ) filament voltage can be set from 5VDC to 13VDC; Low "(Run" ) filament voltage can be set from 5VDC to "Start" setting. Anode Current
Can be preset internally by adjustment of trim pot from 250 to 350 mA. The supply will operate properly with anode voltage drops from 60 to 100VDC with the input line voltage from 90 to 132 VAC rms or 180 to 264 VAC rms. Factory setting is 300.0 mA. Anode Strike Voltage
Inductively generated nominal 700+-50V pk Anode Current Ripple and Noise
Current ripple is less than 0.02%. Long term stability is less than 0.2% after a 30 minute stabilization time. Anode Current Long Term Stability
Current deviation less than 0.3mA after 30 minute stabilizaton Anode Current Thermal Stability
Current deviation less than 0.3 mA from 10 to 40 C degrees deviation measured after 30 minute stabilization. Each unit tested to 60 C degrees for stability. Indicators
AC ON = Green LED RUN = Green LED START UP = Yellow LED Primary FusingPhysical Features
Fused in Both Mains Legs Anode Output Limits
Limited at 0.4 A maximum for Anode to Cathode of Anode to Filament Short Voltage shall be below 60 VDC within 20 seconds after removal of mains power. Warning!
Anode output is high voltage and can cause injury if contacted while unit is running. Filament Output Limits
Thermal Shutoff internal to Regulator IC Limited at 3.5 A maximum for Short Circuit Agency Requirements
CE Marked via TUV Rheinland to EN61010-1 NRTL Listed to IEC and UL950 Passes Class A Conducted EMI tests of TUV and FCC Temperature Limits
Operating = 0 to 40C for maximum relaibility Storage = -20 to 100C Output Connector
Switchcraft DF3 (SC1009) Dimensions
5" high x 5" wide x 8.5" long including mounting lip with Captive screw for quick installation Cooling
Internal cross flow 30CFM, 80x24 mm 5 year life Ball Bearing fan Optional Output Cable
18 " removable cable (Switchcraft R3M from chassis to AMP 350767 Matches popular Lamp connetor